Differential Microcalorimetry of Tracking

microDSC

The Differential Microcalorimetry of Tracking (microDSC) it is especially designed to determine the thermal stability and the heat specific to proteins and other biomacromolecules in diluted dissolutions where the sensitivity of the conventional DSC is not sufficient

Job request

In order to request the analysis of samples for microDSC, you have to get yourselves with the staff responsible for the team in touch.

It be necessary to inform of the toxicity and precautions of storage of the samples. If they do not come to search, the surpluses of sample will be destroyed in the deadline of a week from the delivery of results.

The equipment

MT INSTRUMENTS Nano DSC

Year:
2010



Type of measure of the heat:
Compensation of power
Rank of temperatures:
-10ºC until 130ºC
Speed of warming: :
of 0.0001K/min until 2K/min
Noise:
0,015 μW
Reproductibility of the basis line:
 0,028 μW
Time of answer:
5 s
 Volume of the cell:
0.33 mL

How does it work?

As in the case of the conventional DSC, in the microDSC the heat absorbed or given off is measured by a sample when this undergoes a programme of temperatures. However, the sensitivity of the microDSC is very superior to that of the conventional DSC, making her appropriate for the study of little energetic processes.

At instrumental/operational level, the main ones you differentiate between both techniques they are the way the flow of heat and the type of analysed sample are measured.

Unlike the conventional DSC of the UAT, the team of microDSC works according to the principle of “compensation of power”. The team has two independent ovens, one for the sample and another for the reference. The same programme of temperatures applies to both ovens, but when the sample experiences an exothermic or endothermic process, the moderate temperature will differ from the programmed one. The instrument will regulate the baked applied electrical power of the sample in order to maintain the same temperature in both ovens. This difference of respect power baked of reference will be directly the flow of moderate heat.

Biological samples are analysed by microDSC in dissolution. Unlike the case of the conventional DSC, where the samples are encapsulated, in microDSC the dissolution of sample as well as the reference fill directly in a capillary that acts as cell of measure.

Applications / Practical cases

The microcalorimetry is used widely in the study of biological samples, where the sensitivity of the classical DSC is not sufficient. Among others, it is they can quote the following applications:

• Studies of the union between ligands and proteins
• Characterisation of the stability of biomacromolecules
• Characterisation of the structure of proteins
• Study of lipids and interaction or changes of phase in lipídiques membranes
• Characterisation of nucleic acids, for example transitions B-Z in DNA

More information:

Choose which types of cookies you accept which the University of Girona can store in your browser.

Those that are essential for enabling your connection. There is no option for disabling them, as they are necessary for the functioning of the website.

These enable your options to be remembered (for example language or region you are accessing from), to provide you with advanced services.

They provide statistical information and enable improved services. We use Google Analytics cookies which you can deactivate by installing this plugin.

To offer advertising contents relating to the interests of users, either directly, or through third parties (“adservers”). These must be activated if you wish to see the YouTube videos uploaded to the University of Girona’s website.